DC 10000

DC 10000 dilution cascade with electrical operating pump

Downloads

Brief Description
Brief product description in two-pages as PDF
Product Description
Detailed product description as PDF
DC 10000 freigestellt.png

DC 10000

DC 10000 dilution cascade with electrical operating pump

Description

The DC 10000 consists of four cascaded special dilution systems with the dilution factor 1:10. The DC 10000 dilution cascade has one aerosol inlet and four aerosol outlets. Depending on which aerosol outlet is connected to the measuring device, the aerosol with the dilution factors 1:10, 1:100, 1:1,000 or 1:10,000 are diluted and measured.

The DC 10000 can be operated with all common optical particle counters (OPC) according to ISO 12501-4 or optical aerosol spectrometers (OAS) according to ISO 12501-1. The DC 10000 can be used up to a particle size of approx. 5 µm.

DC 10000 Bild Englisch.JPG

Fig. 1: Overview of the DC 10000, front view

Functional principle

VKL 10 Principle_eng.JPG

Fig. 2: Functional principle

Particle-free air with the volume flow VR circulates through an annular passage around the suction nozzle. Thus, according to Bernoulli, a volume flow VAn is generated in the suction nozzle.

The dilution factor VF is calculated according to the formula:

Verdünnungssysteme Formel.JPG

Fig. 3: Formula for the dilution factor VF.

The DC 10000 needs no compressed air connection. So only an electrical power supply is necessary for operation.

characteristics dilution systems.JPG

Table 1:  Technical characteristics of Palas® dilution systems

Benefits

  • No compressed air connection; only electrical power supply 115 - 230 V, 50 - 60 Hz
  • Dilution factors 1:10, 1:100, 1:1,000, 1:10,000
  • The user can perform a simple functional test on-site
  • The dilution systems can be combined with all common particle counters
  • The dilution systems from Palas® are characterized unambiguously. This is documented with a calibration certificate for each individual device.

Datasheet

Power supply

115 – 230 V, 50 – 60 Hz
Dimensions
approx. 500 • 230 • 150 mm (H • W • D)
Weight
approx. 10 kg
Dilution factor

1 : 10
1 : 100
1 : 1,000
1 : 10,000
Isokinetic suction nozzles 2 – 5 l/min, 15 – 37 l/min
Maximum particle size < 5 µm
Volume flow (clean air)


72 – 180 l/min
Volume flow (suction flow) 2 – 5 l/min

Applications

  • Aerosol dilutions in filter media test systems, e.g. MFP 1000 HEPA and MFP Nano plus according to EN 1822 and ISO 29463
  • Aerosol dilutions in filter test systems, e.g. according to EN 779, ASHRAE 52.2 and ISO 16890
  • Aerosol dilution in clean rooms
  • Aerosol dilution in the operating room to determine the level of protection according to SWKI VA 105-1 and DIN 1946-4
  • Recovery tests according to ISO 14644-3