Promo® 1000 HP

With automatic regulation of sampling volume flow by the aerosol sensors welas® under overpressure up to 10 bar or in temperatures to 120 °C

Downloads

Promo® 1000

Promo® 1000 HP

With automatic regulation of sampling volume flow by the aerosol sensors welas® under overpressure up to 10 bar or in temperatures to 120 °C

Description

Depending on the composition of the aerosol to be measured, i.e. the carrier gas component and the particle material, pressure and temperature changes in the carrier gas can significantly influence the particle size distribution, e.g. due to condensation or evaporation.

For this reason, the aerosol sensor welas ® 1100 HP and the aerosol sensor welas ® 1200 HP are equipped with a cuvette heatable up to 120 °C and pressure-tight up to 10 barg to ensure isobaric and isothermal sampling into the sensor's measurement volume.


The welas® digital is usually calibrated for the operating volume flow. As the operating volume flow changes with pressure and temperature, it is advantageous for the user if automatic volume flow regulation for the sampling volume flow is provided for in the device.

In the Promo® 1000 HP the pressure and temperature of the carrier gas are measured and the required operating volume flow is automatically set to 5 l/min.

Includes:

  • Mass flow controller for volume flow regulation
  • Heating regulator up to 120 °C
  • Temperature sensor
  • Absolute pressure capsule
  • Filter unit

Benefits

  • Messbereich von 200 nm bis 40 µm (3 Messbereiche in einem Gerät wählbar)
  • Bis zu drei Messbereiche in einem Gerät:

    • 0,2 µm – 10 µm
    • 0,3 µm – 17 µm
    • 0,6 µm – 40 µm
  • Größenkanäle bis zu 128 pro Messbereich
  • Konzentrationsbereich von < 1 Partikel/cm3 bis 5 • 105 Partikel/cm3
  • Kalibrierkurven für unterschiedliche Brechungsindizes
  • Sehr hoher und reproduzierbarer Zählwirkungsgrad schon ab 0,2 µm
  • Hohe zeitliche Auflösung von bis zu 10 ms
  • Analyse Software PDAnalyze
  • Kalibrierung, Reinigung und Lampenwechsel können vom Kunden eigenständig durchgeführt werden
  • Externe Ansteuerung über RS 232 oder Ethernet
  • Optional: Software PDControl zum Betrieb als welas® digital erhältlich
  • Einfache Bedienung
  • Wartungsarm
  • Zuverlässige Funktion
  • Senkt Ihre Betriebskosten

Datasheet

Interfaces USB, Ethernet, RS232/485, Wi-Fi
Measurement range (size)


0.2 – 10 µm,
0.3 – 17 µm,
0.6 – 40 µm
Size channels
up to 128 (64/decade)
Measuring principle Optical light-scattering
Measurement range (number CN) < 5 • 105 particles/cm3
Time resolution
up to 1 s
Thermodynamic conditions
10 – 120 °C, 2 – 10 barg
Volume flow 5 l/min, 1.6 l/min
Data acquisition 20 MHz processor, 256 raw data channels, digital
Light source Xenon
Hochdrucklampe 75 W
User interface
Touch screen, 800 • 480 pixel, 7"
Power supply

115 – 230 V, 50 – 60 Hz
Housing Table housing, optionally with mounting brackets for rack-mounting
Dimensions 185 • 450 • 315 mm (H • W • D) (19")
Weight
approx. 8 kg (control unit), 18 kg (sensor)
Operating system

Windows embedded
Data logger storage

4 GB Compact Flash
Software PDControl, FTControl
Installation conditions
+5 – +40 °C (control unit)

Applications

  • Abscheidegradbestimmung von KFZ Innenraumfiltern, Motorluftfiltern, Raumluftfiltern, Druckluftfiltern, Staubsaugerfiltern, abreinigbaren Filtern, Elektrofiltern, Ölabscheidern, Kühlschmierstoffabscheidern, Nassabscheidern, Zyklonen und anderen Abscheidern
  • Isotherme und isobare Partikelgrößen- und Mengenbestimmung, z. B. in der Automobil-, Chemie-, Pharma- und Lebensmittelindustrie
  • Untersuchung schneller, instationärer Prozesse
  • Partikelmessung zur Wolkenbildung
  • Emissionsmessungen

Downloads

Brief Description
Brief product description in two-pages as PDF
Product Description
Detailed product description as PDF

Cookie-Einstellungen

Unsere Webseite verwendet Cookies, um Ihnen ein angenehmeres und sicheres Surfen zu ermöglichen.

Cookies sind für die Funktionalität des Benutzerbereichs sowie für die Warenkörbe notwendig.

Um unser Angebot und unsere Webseite weiter zu verbessern, erfassen wir anonymisierte Daten für Statistiken und Analysen.